Quality Assurance & Standards Flashcards

7 cards from real MLS practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Quality Assurance & Standards flashcards as text
  1. A Levey-Jennings chart shows 6 consecutive control values all falling on the same side of the mean. Which Westgard rule is violated?

    Answer: 10x rule

    The 10x rule is violated when 10 consecutive control values fall on the same side of the mean, indicating systematic error; 6 in a row triggers the warning precursor.

  2. Which term describes the ability of an analytical method to measure only the analyte of interest without interference from other substances?

    Answer: Specificity

    Specificity (or selectivity) refers to the method's ability to measure only the target analyte without cross-reactivity from other compounds.

  3. A laboratory's troponin assay has a coefficient of variation (CV) of 15% at a low concentration. This primarily indicates a problem with:

    Answer: Precision

    CV measures imprecision; a high CV at low concentrations indicates poor reproducibility (precision) of the assay at that level.

  4. CLIA '88 regulations require that proficiency testing samples be tested:

    Answer: By the same method used for patient samples

    CLIA '88 mandates that PT samples be tested using the same routine methods, instruments, and personnel as patient samples to ensure valid assessment.

  5. Which type of error is detected by calculating the mean of control values over time?

    Answer: Systematic error

    Systematic error causes a consistent bias (shift or trend) that is detected by monitoring the mean of control values over multiple runs.

  6. A method comparison study yields a slope of 1.12 and a y-intercept of 0.5. What type of error does the slope indicate?

    Answer: Proportional systematic error

    A slope significantly different from 1.0 indicates proportional systematic error, where the new method reads proportionally higher or lower than the reference.

  7. The delta check in laboratory quality control compares:

    Answer: A patient's current result with their previous result

    The delta check flags results that have changed by more than an expected amount from a patient's previous value, alerting staff to possible specimen mix-up or acute change.