Laboratory Equipment and Quality Questions and Answers Flashcards

6 cards from real MLPAO practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 6 Laboratory Equipment and Quality Questions and Answers flashcards as text
  1. What is the Beer-Lambert Law and how is it applied in clinical chemistry spectrophotometry?

    Answer: It states that absorbance (A) is directly proportional to the concentration of the analyte (c) and path length (b): A = εbc

    The Beer-Lambert Law states A = εbc, where A is absorbance, ε is the molar absorptivity constant, b is path length, and c is concentration. Absorbance increases linearly with analyte concentration.

  2. What is a 'delta check' in laboratory quality management?

    Answer: Comparison of a patient's current result with their previous result to detect unexpected changes

    A delta check compares a patient's current laboratory result with their previous result for the same test to flag unexpected or implausible changes that might indicate specimen mix-up, collection error, or laboratory error.

  3. In hematology, what laboratory parameter is measured by the erythrocyte sedimentation rate (ESR) test?

    Answer: The rate at which red blood cells settle in anticoagulated blood over one hour

    The ESR measures the rate (in mm/hour) at which red blood cells settle in a vertical tube of anticoagulated blood over one hour. It is a non-specific marker of inflammation.

  4. What type of quality control chart is used to monitor the mean and variation of laboratory analytical results over time?

    Answer: Levey-Jennings chart

    The Levey-Jennings chart plots QC results over time against mean and ±1, ±2, and ±3 SD control limits, allowing visual detection of shifts, trends, and random error in laboratory analytical performance.

  5. What is the purpose of the daily performance verification (DPV) for point-of-care testing (POCT) devices?

    Answer: To verify the POCT device performs within acceptable limits using quality control materials before patient testing

    Daily performance verification using QC materials confirms that the POCT device (e.g., glucometer, i-STAT) is functioning correctly before patient results are generated, ensuring result reliability.

  6. What does the coefficient of variation (CV%) measure in laboratory analytical performance?

    Answer: The degree of random variation in a method expressed as a percentage of the mean

    The CV% (coefficient of variation) expresses imprecision (random variation) as a percentage of the mean: CV% = (SD/Mean) × 100. Lower CV% indicates better precision.