  1. What three parameters must be achieved simultaneously for effective steam sterilization?

    Answer: Temperature, time, and steam quality (moisture)

    Steam sterilization requires correct temperature, adequate exposure time, and proper steam quality simultaneously.

  2. What is the difference between high-level disinfection and sterilization?

    Answer: HLD kills all organisms except high numbers of spores; sterilization destroys all forms including spores

    HLD eliminates most organisms but not all spores; sterilization achieves complete microbial destruction.

  3. Why must air be removed from the sterilizer chamber?

    Answer: Air acts as an insulating barrier preventing steam contact, creating cold spots

    Trapped air creates insulating pockets where sterilization temperature is never achieved.

  4. What is the Bowie-Dick test's purpose?

    Answer: To verify vacuum system adequately removes air from the chamber

    The Bowie-Dick test evaluates the air removal capability of prevacuum sterilizers.

  5. What is immediate-use steam sterilization (IUSS) and when is it appropriate?

    Answer: Emergency shortened cycle when no sterile alternative is available

    IUSS is a shortened cycle for emergencies only when no sterile alternative instrument is available.

  6. What is the minimum biological indicator testing frequency for each sterilizer?

    Answer: At least weekly, plus every load with implants

    BIs should be run at least weekly in each sterilizer and with every implant load.