  1. Which sterilization parameter is measured by a Biological Indicator (BI) containing Geobacillus stearothermophilus spores?

    Answer: Steam sterilization efficacy

    G. stearothermophilus spores are the standard BI for steam (autoclave) sterilization validation.

  2. What is the primary mechanism by which gamma radiation achieves sterilization?

    Answer: DNA strand breaks and free radical generation

    Gamma radiation kills microorganisms by inducing DNA damage through direct ionization and reactive free radicals.

  3. A tissue bank uses electron beam (e-beam) sterilization. Which feature distinguishes e-beam from gamma radiation?

    Answer: Much faster processing time

    E-beam processing is significantly faster than gamma because it uses accelerated electrons rather than isotope decay.

  4. During ETO sterilization, which post-processing step is mandatory before tissue release?

    Answer: Aeration to remove ETO residuals

    Aeration is required after ETO sterilization to reduce toxic residual ethylene oxide to safe levels.

  5. A Sterility Assurance Level (SAL) of 10⁻⁶ means:

    Answer: One in a million units may harbor a viable microorganism

    SAL 10⁻⁶ means there is no more than a 1-in-1,000,000 probability that a single viable microorganism remains on a sterilized item.

  6. Which tissue type is MOST sensitive to gamma radiation-induced collagen degradation?

    Answer: Soft tissue tendons and ligaments

    Soft tissue grafts like tendons and ligaments are particularly vulnerable to radiation-induced collagen cross-linking and structural weakening.

  7. What is the purpose of a dose-mapping study in radiation sterilization?

    Answer: To identify minimum and maximum dose points within a product load

    Dose mapping identifies the minimum and maximum radiation dose locations within the product configuration to ensure the entire load receives a sterilizing dose.