Load Cell Wiring & Troubleshooting Flashcards

7 cards from real CST practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Load Cell Wiring & Troubleshooting flashcards as text
  1. What is the typical color code for the excitation positive (+EXC) wire in a standard 4-wire load cell cable?

    Answer: Red

    In most standard load cell color conventions, red is used for the excitation positive (+EXC) lead.

  2. A technician measures 0 mV output from a load cell under full rated load. Which fault is most likely?

    Answer: Open excitation circuit

    An open excitation circuit removes the bridge supply voltage, resulting in zero output regardless of applied load.

  3. When wiring multiple load cells in a parallel junction box, what must be matched between cells to ensure accurate summing?

    Answer: Input impedance of each load cell

    Matching input impedance ensures equal current distribution across parallel-connected load cells for accurate summing.

  4. What does a negative mV/V output reading indicate when a load cell is under a compressive load that should produce a positive reading?

    Answer: Signal wires are reversed

    A negative output when positive is expected means the signal+ and signal− leads have been swapped.

  5. Which instrument is most appropriate for measuring the zero balance output (offset) of a load cell in mV?

    Answer: Precision digital multimeter

    A precision digital multimeter with mV resolution is the correct tool for measuring the small DC offset voltage of a load cell at zero balance.

  6. What wiring configuration is used in a 6-wire load cell to compensate for cable resistance errors?

    Answer: Separate sense lines feed back excitation voltage to the indicator

    6-wire load cells add sense lines that allow the indicator to monitor and regulate the exact voltage at the load cell terminals, eliminating cable resistance drop errors.

  7. A junction box trim potentiometer is adjusted to correct for what specific condition in a multi-load-cell system?

    Answer: Corner or span errors between individual load cells

    Trim potentiometers in a summing junction box allow individual cell output levels to be equalized, correcting corner and span errors.