Troubleshooting Techniques & Repair Methods Flashcards

7 cards from real CST practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Troubleshooting Techniques & Repair Methods flashcards as text
  1. When using a summing card (junction box) to diagnose individual load cells, what measurement confirms a shorted load cell?

    Answer: Lower than specified bridge resistance

    A shorted load cell will exhibit lower than specified bridge resistance because additional current paths reduce the effective resistance.

  2. A platform scale indicator shows 'ERR-2' and will not accept a zero command. According to typical manufacturer conventions, what does this type of error most likely indicate?

    Answer: Weight is outside the acceptable zero range

    Most scale indicators use error codes like 'ERR-2' or similar to indicate that the current weight is outside the permissible zero band for the zero command.

  3. A scale technician notices that a scale reads correctly in the morning but reads high by several graduations in the afternoon. What phenomenon is most likely responsible?

    Answer: Thermal expansion of the platform structure causing mechanical stress on load cells

    Thermal expansion of the platform during the warmer part of the day can induce mechanical stress on load cells, causing artificially elevated readings.

  4. When replacing a sealed load cell, a technician must apply sealant around the cable entry point. What is the primary purpose of this step?

    Answer: To prevent moisture from wicking into the load cell along the cable

    Sealing the cable entry prevents moisture from entering the load cell via capillary action along the cable, which would degrade insulation resistance and cause signal errors.

  5. A retail price-computing scale fails its shift test (constant load test). The reading drifts more than the permissible tolerance over 30 minutes. Which component is most likely to blame?

    Answer: A load cell exhibiting excessive creep

    The shift test measures reading stability over time under a constant load; excessive creep in a load cell causes the output to slowly drift and fail this test.

  6. A technician finds that a scale's output increases when the load cell cable is flexed near the junction box. What does this symptom indicate?

    Answer: An intermittent open or high-resistance connection in the cable

    An intermittent open or high-resistance connection causes the signal to change when the cable is physically disturbed, confirming a wiring fault at or near that location.

  7. A crane scale consistently under-reads by a fixed amount regardless of the load applied. Zero is correct with no load. What is the most probable cause?

    Answer: Incorrect deadweight or tare configuration in the indicator

    A fixed absolute under-read at all loads while zero is correct indicates an incorrectly programmed tare or deadweight value subtracting from the gross reading.