Sterilization Methods Flashcards
7 cards from real CRCST practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.
Read the first 7 Sterilization Methods flashcards as text
What is the purpose of the Bowie-Dick test in steam sterilization?
Answer: To verify air removal and steam penetration in prevacuum sterilizers
The Bowie-Dick test verifies adequate air removal and uniform steam penetration in prevacuum (dynamic air removal) steam sterilizers.
Low-temperature sterilization using ozone works primarily through which mechanism?
Answer: Oxidation of cellular components
Ozone sterilization destroys microorganisms through oxidation of cellular components including proteins, lipids, and nucleic acids.
The sterilization assurance level (SAL) for medical devices is defined as a probability of not more than:
Answer: 1 in 1,000,000
The SAL for sterile medical devices is 10⁻⁶, meaning no more than a 1 in 1,000,000 chance of a viable microorganism remaining.
Which type of chemical indicator changes color based on exposure to specific sterilization parameters?
Answer: Class 5 — integrating indicators
Class 5 integrating indicators react to all critical parameters (time, temperature, and sterilant) and are designed to correlate with biological indicator performance.
What minimum aeration time is typically required after EtO sterilization at ambient room temperature (20°C)?
Answer: 7 days
At room temperature (20°C), EtO-sterilized items require approximately 7 days of aeration to reduce residual EtO to safe levels.
Which statement correctly describes the difference between sterilization and disinfection?
Answer: Sterilization destroys all forms of microbial life; disinfection reduces but may not eliminate all microorganisms
Sterilization destroys all forms of microbial life including spores, while disinfection reduces microbial counts but may not eliminate all forms.
Formaldehyde-based low-temperature sterilization is classified as which level of germicide?
Answer: Sporicide and sterilant
Formaldehyde at appropriate concentrations and exposure times is classified as a sporicide and sterilant capable of achieving sterilization.