โ† All CMT Flashcard Decks

Monitoring Techniques and Equipment Flashcards

7 cards from real CMT practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 Monitoring Techniques and Equipment flashcards as text
  1. Which monitoring lead is considered best for detecting ST-segment changes during ischemia monitoring?

    Answer: Lead V5 (or MCL5)

    Lead V5 (or MCL5) is most sensitive for detecting lateral wall ischemia and ST-segment changes, making it preferred for ischemia monitoring.

  2. What is the correct skin preparation technique before applying monitoring electrodes to maximize signal quality?

    Answer: Clean skin with alcohol, allow to dry, and lightly abrade if needed

    Skin should be cleaned with alcohol, allowed to dry completely, and lightly abraded to remove dead skin cells and reduce impedance.

  3. A telemetry transmitter battery is low. What is the most likely effect on monitoring quality?

    Answer: Signal dropout, artifact, or loss of transmission

    Low battery in a telemetry transmitter causes signal weakness, resulting in signal dropout, artifact, or complete loss of transmission to the monitoring station.

  4. Which factor most directly affects the radiofrequency signal range of a telemetry system?

    Answer: Building construction materials and competing wireless signals

    Building materials (concrete, steel) and competing wireless devices (Wi-Fi, other telemetry) are primary factors limiting telemetry transmission range.

  5. What is the function of the RA electrode in a standard 3-lead monitoring setup?

    Answer: It serves as the negative electrode and reference for Lead II

    In Lead II, the RA electrode serves as the negative pole, while the LL electrode is positive, with the LA electrode serving as ground.

  6. When using a 5-lead monitoring system compared to a 3-lead system, what additional capability is provided?

    Answer: Simultaneous monitoring of a precordial chest lead (e.g., V1)

    A 5-lead system adds a chest (precordial) electrode, allowing simultaneous display of limb leads and one chest lead for enhanced diagnostic capability.

  7. A patient's telemetry shows intermittent artifact that occurs every time the patient raises their left arm. What is the most likely cause?

    Answer: Lead wire or electrode on the left arm or shoulder becoming intermittently loose

    Motion-related artifact that correlates with arm movement indicates a loose electrode or lead wire at the left arm/shoulder attachment site.