← All CET Flashcard Decks

Echocardiography Imaging Techniques Flashcards

7 cards from real CET practice questions. Tap to flip, then mark Knew It or Still Learning — missed cards come back until you master them.

Read the first 7 Echocardiography Imaging Techniques flashcards as text
  1. Which transducer frequency range is most commonly used for adult transthoracic echocardiography?

    Answer: 2-4 MHz

    Adult TTE typically uses 2-4 MHz transducers to balance penetration depth with adequate resolution for cardiac structures.

  2. In tissue harmonic imaging, the ultrasound system processes which returning frequency to create the image?

    Answer: Second harmonic (2× transmit frequency)

    Tissue harmonic imaging uses the second harmonic frequency generated by nonlinear propagation through tissue, improving signal-to-noise ratio and reducing artifacts.

  3. When imaging a patient with a poor acoustic window, which maneuver can improve transthoracic image quality?

    Answer: Positioning the patient in the left lateral decubitus position

    Left lateral decubitus positioning moves the heart closer to the chest wall and displaces lung tissue, substantially improving acoustic access.

  4. Which imaging mode provides a single-dimension, time-motion recording of cardiac structures and is ideal for precise timing of events?

    Answer: M-mode

    M-mode (motion mode) displays a single line of information over time, offering excellent temporal resolution for measuring wall motion and valve timing.

  5. The acoustic shadow artifact seen posterior to a heavily calcified aortic valve is caused by:

    Answer: Complete attenuation of the ultrasound beam by the dense structure

    Calcified structures completely attenuate the ultrasound beam, preventing sound from reaching structures beyond, creating an acoustic shadow.

  6. Which echocardiographic view is obtained with the transducer placed at the subxiphoid position and is particularly useful when other windows are limited?

    Answer: Subcostal four-chamber

    The subcostal (subxiphoid) four-chamber view uses the liver as an acoustic window and is valuable in patients with COPD, obesity, or post-surgical chests.

  7. In 2D echocardiography, increasing the imaging depth setting will:

    Answer: Decrease frame rate

    Increasing depth requires the system to wait longer for returning echoes before firing the next pulse, reducing the number of frames displayed per second.