A calibration lab receives a flow meter that uses a turbine rotor. Which calibration method provides the most accurate volumetric flow reference?
Answer: Gravimetric method using a weigh tank and timer
The gravimetric method—collecting fluid into a weigh tank over a timed interval—provides a primary-level flow reference by directly measuring mass and time.
When using a coordinate measuring machine (CMM) to calibrate a precision fixture, the stylus qualification step is performed to:
Answer: Determine the exact center and effective radius of the stylus ball
Stylus qualification (probing a reference sphere) determines the stylus ball's effective center position and radius, which the CMM software uses to correct all subsequent measurements.
A spectrum analyzer is used during RF power amplifier calibration. The instrument's resolution bandwidth (RBW) setting primarily affects:
Answer: The ability to resolve closely spaced signals and the noise floor
Narrower RBW improves frequency resolution between adjacent signals and lowers the displayed noise floor, while wider RBW speeds sweep time but reduces selectivity.
Which technique is used to verify the linearity of a torque sensor over its full measurement range during calibration?
Answer: Apply certified torque at multiple ascending points and calculate deviation from best-fit line
Linearity is evaluated by applying known torques at multiple points across the range and comparing the actual output to the ideal straight-line response.
A precision resistance standard is kept in a temperature-controlled oil bath during calibration. The primary reason for this practice is:
Answer: To reduce the effect of temperature coefficient of resistance on the standard's value
Precision resistance standards have temperature coefficients that cause their resistance to change with temperature; oil bath control minimizes this variation during calibration.
When calibrating a sound level meter, the calibrator is placed over the microphone and the meter reading is compared to the calibrator's certified output level. If the calibrator produces 94 dB SPL and the meter reads 93.4 dB, the meter has an error of:
Answer: -0.6 dB
The meter reads 0.6 dB below the calibrator's true 94 dB output, so the meter error is -0.6 dB.
Which statement correctly describes the difference between repeatability and reproducibility in the context of calibration equipment performance?
Answer: Repeatability is variation under the same conditions by the same operator; reproducibility includes variation across operators, setups, or time
Repeatability quantifies variation under fixed conditions (same operator, instrument, environment), while reproducibility captures variation when one or more of those conditions changes.