What is the primary purpose of the Total Cell Volume (TCV) or Fiber Bundle Volume (FBV) test during dialyzer reprocessing?
Answer: To ensure the dialyzer can still effectively clear waste products
The TCV test measures the internal volume of the hollow fibers. A significant drop in TCV, typically below 80% of the original volume, indicates that clotting or protein deposits have blocked too many fibers, reducing the dialyzer's surface area and its ability to provide adequate clearance of waste products.
Which chemical is most commonly used as the sterilant/disinfectant in modern dialyzer reprocessing systems?
Answer: Peracetic Acid
Peracetic acid (PAA) is a widely used high-level disinfectant and sterilant in dialyzer reprocessing. It is effective against a broad spectrum of microorganisms, including bacteria, viruses, and spores, and it breaks down into non-toxic byproducts (acetic acid and oxygen).
Before a reprocessed dialyzer is used for a patient's treatment, what is the most critical final check that must be performed at the patient's station?
Answer: Testing for residual sterilant
It is absolutely critical to ensure that no residual sterilant remains in the dialyzer before it is connected to the patient. Exposure to even small amounts of chemicals like peracetic acid can cause serious adverse reactions, including hemolysis and anaphylaxis. This is tested for right before use.
According to AAMI standards, a reprocessed dialyzer must be discarded if its fiber bundle volume (FBV) drops below what percentage of its original measured volume?
Answer: 80%
The Association for the Advancement of Medical Instrumentation (AAMI) sets the standard that a reprocessed dialyzer must retain at least 80% of its original TCV/FBV. If the volume drops below this threshold, the dialyzer has lost too much surface area to provide an adequate treatment and must be discarded.
What does a failed pressure test during dialyzer reprocessing indicate?
Answer: A leak in the dialyzer membrane
A pressure test, or leak test, checks the integrity of the hollow fiber membrane. If the dialyzer cannot hold a set pressure, it signifies a leak, which could allow non-sterile dialysate to mix with the patient's blood, posing a significant risk of infection and blood loss.
Proper labeling of a reprocessed dialyzer is crucial for patient safety. Which of the following pieces of information is required on the label?
Answer: The number of previous uses
The label on a reprocessed dialyzer must contain critical patient safety information to ensure it is used on the correct patient and is still safe for use. This includes the patient's name, the number of times it has been used, and the date of reprocessing.
During the reprocessing procedure, what is the purpose of the 'reverse ultrafiltration' step?
Answer: To force cleaning agents from the dialysate compartment to the blood compartment
Reverse ultrafiltration pushes fluid (water and cleaning agents) in the opposite direction of normal dialysis flow, from the dialysate compartment into and through the blood compartment. This helps to dislodge and flush out protein deposits and blood residue from the inner surface of the hollow fiber membranes, ensuring a more thorough cleaning.