General Practice Flashcards

7 cards from real CBET practice questions. Tap to flip, then mark Knew It or Still Learning โ€” missed cards come back until you master them.

Read the first 7 General Practice flashcards as text
  1. A pulse oximeter reads 98% SpO2 but the patient appears cyanotic. Which condition MOST likely explains this discrepancy?

    Answer: Carbon monoxide poisoning causing carboxyhemoglobin

    Pulse oximetry cannot distinguish carboxyhemoglobin from oxyhemoglobin, so CO poisoning produces falsely elevated SpO2 readings despite true hypoxia.

  2. The primary purpose of a Faraday cage in a medical imaging suite is to:

    Answer: Prevent external RF signals from interfering with MRI signals

    A Faraday cage blocks external radiofrequency signals that would interfere with the extremely weak MRI signals being detected.

  3. When calibrating a patient scale used for medication dosing, a biomedical technician should use NIST-traceable weights primarily to:

    Answer: Ensure measurements are accurate relative to a national standard

    NIST-traceable weights provide an unbroken chain of calibration to national standards, ensuring measurement accuracy and regulatory compliance.

  4. In a 3-phase electrical system supplying medical imaging equipment, an open neutral conductor would MOST likely cause:

    Answer: Overvoltage on some phases damaging connected equipment

    An open neutral in a 3-phase system causes voltage imbalance, resulting in overvoltage on some phases that can damage connected equipment.

  5. A surgical electrosurgical unit (ESU) causes interference on a nearby patient monitor. The FIRST corrective action should be:

    Answer: Check and ensure proper placement and contact of the ESU dispersive electrode

    Improper dispersive electrode placement causes high-frequency current to seek alternate paths, creating interference in nearby monitoring equipment.

  6. Which parameter is checked during an acceptance test for a defibrillator to verify energy delivery accuracy?

    Answer: Delivered energy measured with a calibrated defibrillator analyzer

    A calibrated defibrillator analyzer measures actual delivered energy into a known load, verifying the device delivers the energy it displays.

  7. Which law governs the relationship between pressure, volume, and temperature for an ideal gas, relevant to medical gas cylinders?

    Answer: The Combined Gas Law (Boyle's + Charles' Laws)

    The Combined Gas Law (P1V1/T1 = P2V2/T2) describes how pressure, volume, and temperature interrelate in gas cylinders used in medical settings.