  1. What is the primary difference between a Class I and Class II medical device under FDA classification?

    Answer: Class I requires general controls only; Class II requires special controls such as performance standards

    FDA Class I devices are subject to general controls only, while Class II devices require special controls (e.g., performance standards, post-market surveillance) to ensure safety and effectiveness.

  2. A physiological pressure transducer requires zeroing before use. Why is this step necessary?

    Answer: To establish atmospheric pressure as the reference baseline, eliminating hydrostatic and offset errors

    Zeroing the transducer at the level of the right atrium sets atmospheric pressure as the zero reference, ensuring accurate hemodynamic readings.

  3. Which parameter does a capnograph measure, and what waveform phase indicates end-tidal CO2?

    Answer: CO2 concentration in exhaled breath; the alveolar plateau (Phase III)

    Capnography measures exhaled CO2, and the end-tidal CO2 (EtCO2) value is read at the alveolar plateau (Phase III) of the capnogram waveform.

  4. During a PM on a hospital bed, you discover the brake does not engage fully. What is the correct action?

    Answer: Tag the bed out-of-service and remove from patient use pending repair

    A bed with a failed brake poses a fall and injury risk; it must be immediately tagged out-of-service and removed from clinical use until repaired and verified.

  5. What is the purpose of a Faraday cage in an MRI suite?

    Answer: To shield external RF signals that would corrupt MRI image quality

    The RF-shielded Faraday cage prevents external radiofrequency interference from entering the MRI room and corrupting the weak MR signal during image acquisition.

  6. A hospital is evaluating a new ultrasound machine. Under FDA guidance, which testing verifies the acoustic output does not exceed safe limits?

    Answer: AIUM/NEMA measurement standards for acoustic output (MI and TI indices)

    Mechanical Index (MI) and Thermal Index (TI) are standardized acoustic output metrics per AIUM/NEMA guidelines that the FDA uses to regulate safe ultrasound exposure levels.

  7. A nurse reports that an automated external defibrillator (AED) failed to advise a shock during a cardiac arrest drill. After replacing the dummy pad, it still shows 'check electrodes.' What should the BMET do next?

    Answer: Test the AED's electrode impedance sensing circuit using a calibrated AED analyzer

    A calibrated AED analyzer can simulate patient electrode impedance to verify whether the AED's electrode sensing and analysis circuits are functioning correctly.