  1. What is the standard sterilization temperature and time for a gravity displacement steam autoclave at 121°C?

    Answer: 15 minutes

    Gravity displacement autoclaves operating at 121°C (250°F) require a minimum of 15 minutes at that temperature to achieve sterilization.

  2. Which Spaulding classification category requires sterilization rather than just high-level disinfection?

    Answer: Critical items

    Critical items enter sterile tissue or the vascular system and must be sterilized to eliminate all microorganisms including spores.

  3. What is the primary biological indicator used to validate steam sterilization processes?

    Answer: Geobacillus stearothermophilus spores

    Geobacillus stearothermophilus is the most resistant organism to moist heat and is the standard biological indicator for steam sterilization validation.

  4. A BIOMED technician is testing an EtO (ethylene oxide) sterilizer. Which parameter is NOT critical to monitor during the EtO cycle?

    Answer: UV light exposure

    EtO sterilization depends on gas concentration, temperature, humidity, and exposure time — UV light plays no role in this chemical sterilization process.

  5. What does a positive result on a Bowie-Dick test indicate?

    Answer: Air removal failure in the pre-vacuum autoclave

    A positive (failed) Bowie-Dick test indicates that air was not properly removed from the chamber, which would prevent steam penetration and sterilization.

  6. Which sterilization method is most appropriate for heat-sensitive endoscopes?

    Answer: Low-temperature hydrogen peroxide plasma

    Low-temperature hydrogen peroxide gas plasma sterilization operates below 50°C and is compatible with heat-sensitive devices like flexible endoscopes.

  7. When performing preventive maintenance on an autoclave, what should the BIOMED technician check first after a reported cycle abort?

    Answer: The printout or cycle log for the fault code

    The cycle log or printout provides the specific fault code and the point of failure, which is the most efficient starting point for diagnosing an autoclave abort.