COT Ophthalmic Instrumentation and Equipment Maintenance — Questions and Answers
Question 1: A Goldmann applanation tonometer is being calibrated using the calibration bar. At the 20 mmHg position, a reading of 22 mmHg is obtained. What action should be taken?
- Continue using the instrument; a 2 mmHg deviation is within acceptable limits
- Remove the instrument from service and schedule repair (Correct answer)
- Re-zero the tonometer and repeat the measurement only at 0 mmHg
- Switch to a non-contact tonometer for all patients that day
Correct answer: Remove the instrument from service and schedule repair
The acceptable calibration tolerance for the Goldmann applanation tonometer is ±0.5 mmHg. A deviation of 2 mmHg at the 20 mmHg check point exceeds this tolerance, meaning the instrument must be taken out of service and sent for calibration or repair before further clinical use.
Question 2: Which disinfection method is recommended between patients for a Goldmann applanation tonometer prism to prevent transmission of adenovirus and other pathogens?
- Autoclave sterilization at 121°C for 15 minutes
- Soak in 1:10 diluted bleach for 30 seconds
- Wipe with 70% isopropyl alcohol and allow to air dry for at least 5 minutes (Correct answer)
- Rinse with sterile saline and dry with gauze
Correct answer: Wipe with 70% isopropyl alcohol and allow to air dry for at least 5 minutes
The CDC and JCAHPO recommend disinfecting tonometer tips between patients by wiping with 70% isopropyl alcohol and allowing a contact time of at least 5 minutes before reuse. Autoclaving damages the prism coating, and saline alone does not adequately disinfect.
Question 3: A keratometer is producing mire images that appear distorted and asymmetric even after lens cleaning. The MOST appropriate next step is to:
- Perform the measurement anyway and document the irregularity in the chart
- Verify accuracy by measuring a steel calibration sphere of known radius (Correct answer)
- Replace the eyepiece lens immediately
- Use the autorefractor reading as a substitute
Correct answer: Verify accuracy by measuring a steel calibration sphere of known radius
When a keratometer produces suspect or distorted readings, the standard troubleshooting step is to measure a steel calibration sphere of known radius to verify whether the instrument is producing accurate values. This confirms whether the issue is instrument calibration or corneal irregularity.
Question 4: When cleaning the objective lens of a slit lamp biomicroscope, the correct technique is to:
- Use a dry cotton swab in a vigorous circular motion to remove debris
- Apply 70% isopropyl alcohol directly to the glass surface with a paper towel
- Use a lens-cleaning solution with a lint-free cloth, wiping gently from center outward (Correct answer)
- Rinse with tap water and pat dry with a soft cloth
Correct answer: Use a lens-cleaning solution with a lint-free cloth, wiping gently from center outward
Optical surfaces should be cleaned using an appropriate lens-cleaning solution and a lint-free cloth, wiping from center outward to avoid dragging debris across the optical surface. Isopropyl alcohol can damage coatings, and dry or abrasive materials can scratch the lens.
Question 5: Which sterilization method is MOST appropriate for ophthalmic surgical instruments with heat-sensitive components?
- Steam autoclave at 134°C
- Dry heat oven at 160°C for 2 hours
- Ethylene oxide (EtO) gas sterilization (Correct answer)
- Boiling in distilled water for 10 minutes
Correct answer: Ethylene oxide (EtO) gas sterilization
Ethylene oxide (EtO) gas sterilization achieves full sterilization at low temperatures (37–63°C), making it safe for heat-sensitive materials such as plastics, rubber, and certain instrument components that would be damaged by autoclaving or dry heat. Boiling does not achieve sterilization.
Question 6: During routine maintenance checks, which action should be performed DAILY on a non-contact tonometer (NCT)?
- Replace the internal air puff mechanism filter
- Perform a calibration check using the instrument's built-in verification procedure (Correct answer)
- Disassemble and clean the nozzle tip in an autoclave
- Lubricate the joystick mechanism with medical-grade oil
Correct answer: Perform a calibration check using the instrument's built-in verification procedure
Non-contact tonometers should undergo a daily calibration verification check using the instrument's built-in self-calibration or a reference target to ensure accurate IOP readings. Nozzle tips are wiped with alcohol between patients, not autoclaved, and internal components are serviced by biomedical technicians.
A Goldmann applanation tonometer is being calibrated using the calibration bar.
At the 20 mmHg position, a reading of 22 mmHg is obtained.
What action should be taken?