Certified Hemodialysis Technologist/Technician Exam Water Treatment and Quality 2 — Questions and Answers
Question 1: What is the AAMI maximum allowable bacterial count in dialysate?
- 100 CFU/mL
- 200 CFU/mL (Correct answer)
- 500 CFU/mL
- 1000 CFU/mL
Correct answer: 200 CFU/mL
AAMI standards allow a maximum of 200 CFU/mL of bacteria in dialysate.
Question 2: Activated carbon tanks in a water treatment system primarily remove which contaminants?
- Heavy metals
- Chlorine and chloramines (Correct answer)
- Hardness ions
- Nitrates
Correct answer: Chlorine and chloramines
Activated carbon adsorbs chlorine and chloramine from the feed water, protecting downstream components.
Question 3: A patient develops sudden hemolysis during dialysis. Which water contaminant should be suspected first?
- Fluoride
- Nitrate
- Chloramine (Correct answer)
- Aluminum
Correct answer: Chloramine
Chloramine exposure during dialysis directly damages red blood cell membranes, causing acute hemolysis.
Question 4: How frequently should the deionizer resin tanks be tested for resistivity in a dialysis water system?
- Each treatment shift (Correct answer)
- Daily
- Weekly
- Monthly
Correct answer: Each treatment shift
Deionizer resin exhaustion must be monitored with resistivity testing at each treatment shift to ensure ion removal effectiveness.
Question 5: What minimum water-to-concentrate ratio is used by most proportioning hemodialysis machines?
- 30:1
- 34:1 (Correct answer)
- 36:1
- 40:1
Correct answer: 34:1
Most proportioning systems dilute bicarbonate or acetate concentrate at a 34:1 (water to concentrate) ratio to create dialysate.
Question 6: Which method is used to disinfect dialysis water distribution loops?
- UV radiation only
- Peracetic acid or bleach recirculation (Correct answer)
- Ozone exposure only
- Heat pasteurization only
Correct answer: Peracetic acid or bleach recirculation
Chemical disinfection with peracetic acid or sodium hypochlorite (bleach) is the most common method for disinfecting dialysis water loops.
What is the AAMI maximum allowable bacterial count in dialysate?